Detail of Recall

Type of Product:Device
Medsafe Reference:19664
Brand Name:GE Drive Gas Check Valve on Anaesthesia Work Stations
Model:Aespire 7900, Aespire View 7900, Aestiva MRI, Aisys, Aisys CS2, Avance and Avance CS2 installed from April 20 2015 through to Oct 2015.
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Brand Name:GE Service Kits (containing the Drive Gas Check) for Anaesthesia Work Stations
Model:1009-8216-000, 1503-3006-000, 1503-8102-0000, 1009-8423-000 and 1503-8101-000.
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The Drive Gas Check could become stuck in a fixed open position which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved it could result n excessive or prolonged pressure in the patient breathing circuit during ventilatoin potentially resulting in barotrauma.There have been no injuries reported as a result of this issue.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/1/2016