Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19664 |
| Brand Name: | GE Drive Gas Check Valve on Anaesthesia Work Stations |
| Model: | Aespire 7900, Aespire View 7900, Aestiva MRI, Aisys, Aisys CS2, Avance and Avance CS2 installed from April 20 2015 through to Oct 2015. |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Datex-Ohmeda Inc |
| Brand Name: | GE Service Kits (containing the Drive Gas Check) for Anaesthesia Work Stations |
| Model: | 1009-8216-000, 1503-3006-000, 1503-8102-0000, 1009-8423-000 and 1503-8101-000. |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Datex-Ohmeda Inc |
| Issue: | The Drive Gas Check could become stuck in a fixed open position which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved it could result n excessive or prolonged pressure in the patient breathing circuit during ventilatoin potentially resulting in barotrauma.There have been no injuries reported as a result of this issue. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/1/2016 |