Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19663
Brand Name:Beckman Coulter iChem VELOCITY Automated Urine Chemistry System using iChem VELOCITY Urine Chemistry Strips
Model:800-7204, 800-7204FJR
Affected:all lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:
Issue:Customer complaints have been received indicatiing loose or missing analyte pads for the iChem VELOCITY Urine Chemistry strips. This may result in false negative results for any analyte(s) or false low or false high pH, but does not apply to the color, clarity and specific gravity which is analyzed through a different mechanism and module. Hazardous situation is a potential erroneous result received by a physician.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/1/2016