Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19644
Brand Name:HeartWare Vetricular Assist Device AC Adapters and batteries
Model:
Affected:
Software version:
Recalling Organisation:Health New Zealand - Te Whatu Ora
9 Ruskin Street
Parnell
Auckland 1052
Contact Information:
Manufacturer:HeartWare International Inc
Issue:As part of HeartWare's ongoing product performance monitoring, we have reviewed certain complaints related to the HVAD® System and are distributing this notice to announce (i) a voluntary recall of AC Adaptors designed for 220-volt power grids; and (ii) a voluntary recall of certain older batteries more likely to exhibit deterioration of battery capacity. HeartWare believes that these actions will help reduce the occurrence of avoidable injuries as patients remain on the device for increasingly long periods of time.
New Field Action - DEC2015A (HVAD Controller AC Adaptors):
o HeartWare FSCA DEC2015A is assigned to product Code: 1430XX; Serial Numbers: CAC000001 to CAC005796; Reason: HeartWare AC Adapters are vulnerable to power grid surges above 220 volts.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/1/2016
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /