Detail of Recall

Type of Product:Device
Medsafe Reference:19634
Brand Name:Ortho-Clinical Diagnostics Vitros 350 Chemistry Systems
Model:6802153
Affected:10758750002054
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics c/- Johnson & Johnson
13A Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 800 705
Manufacturer:Ortho-Clinical Diagnostics Inc
Issue:Manufacturer has initiated a recall due to the need to clarify operator actions folloing U90-382 or 6LU cnditions codes generated by the system.
If a U90-382 or 6LU condition code (i.e associated with a wash error) occurs, the condition code text located on the VITRO System and other user documentation indicates to dilute the sample. However dilution may not be the appropriate action for all scenarios.
This notification provides additional information to a Product Correction Notification that Ortho Clinical Diagnostics (Ortho) issued in December 2015 (Medsafe Ref. 19634) regarding the clarification of operator actions for wash errors. Since that time, Ortho has been monitoring the occurrence of U90-382 or 6LU condition codes associated with wash errors. We observed an increase in wash errors for VITROS Chemistry Products PHYT Slides.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/11/2016