Detail of Recall

Type of Product:Device
Medsafe Reference:19628
Brand Name:Radiometer AQURE Data Management System
Model:
Affected:
Software version:all software versions below V2.1.1
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:Manufacturer has recently become aware of a potential clinical issue with AQURE if the Data Management system is being used to interpret patient results for the management of treatment.
In AQURE, there are two views called Flexlink and Patient View. Each can display multiple patient result values from a single patient, provided over time, from one or multiple devices. This screen is intended to be reviewed by a clinician.
The issue occurs when a sample type is not provided from a device. An example is where results from a sample type of “Urine” (not displayed) is compared to results from a sample type “Arterial”. In this case a clinician may incorrectly take action based on an assumption that all results are from a sample type “Arterial”.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/1/2016