Detail of Recall

Type of Product:Device
Medsafe Reference:19605
Brand Name:Siemens MAGNETOM Systems
Model:multiple numbers
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Brand Name:Siemens MR - MAGNETOM Concerto
Model:
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Brand Name:Siemens - _Magnetom Permanent
Model:
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Shenzhen Magnetic Resonance Ltd
Brand Name:Siemens MRI system
Model:
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Shenzhen Magnetic Resonance Ltd
Brand Name:Siemens Biograph mMR PET/MRI
Model:10433372
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:The manufacture confirms that in some cases the quench lines for some superconducting magnets may have not been properly installed and wants to alert the users to check if the quench lines for their systems have been installed under safe conditions. Siemens in New Zealand will inspect all quench lines installed by Siemens.
Quench lines are installed to vent helium gas in the event of a quench. (Quenching refers to the events that occur when the liquid cryogens that cool the magnet coils boil off rapidly, which results in helium escaping very rapidly from the cryogen bath) In the rare event of a quench an improperly installed quench line may cause the helium gas to be blown directly into the magnet room or other areas.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/12/2015