Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19603
Brand Name:BD Vacutainer Heparin Tube
Model:367885
Affected:Lot No: 5084266
Software version:
Recalling Organisation:Becton Dickinson Ltd
Unit 4 / 110 Mays Road
Onehunga
Auckland 1061
Contact Information:0800 572 468
Manufacturer:Becton Dickinson UK Ltd
Issue:Manufacturer has received a customer complaint advising that have found tubes containing a white powder. One of these tubes had been used on a patient to measure troponin levels, generating a zero result. Patient retested using another tube that did not contain the white powder and different result obtained.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:
Date Commenced:23/12/2015
Hide menus