Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19578
Brand Name:Philips IntelliVue Information Centre (PIIC)
Model:866023 Intellivue Info Center iX, 866117 PIIC Classc Upgrade and 866389 IntelliVue Info Center iX
Affected:all versions
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Calcualated leads, also referred to as reconstructed ECG leads viewed or printed at the Information Centre iX may misrepresent the ECG waveform in specific leads.
Lead combinations that are affected:
Lead 1 and Lead 111 are used to derive the aVL Lead (incorrectly derived)
Lead 1 and Lead aVR are used to derive the aVL Lead (incorrecty derived)'
Lead 111 and Lead aVF are used to derive Lead 11 (incorrecty derived)
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/12/2015
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /