Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19557 |
| Brand Name: | Elekta Mosaiq 2.5 |
| Model: | Mosaiq |
| Affected: | |
| Software version: | 2.5 and higher |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Impac Medical Systems Inc |
| Issue: | Incorrect drug dosage due to "Age Limit" data item issue: When there is no valid (unexpired) patient weight entered on the same date or prior to a serum creatinine value being populated, a stale Creatinine Clearance (CrCl) value is being used in Area under the Curve (AUC) dose calculations. Incorrect Drug Dosage due to patient weight data item issue: If there is a valid age limit added to an Elekta metric data item (eg; weight (kg)) and the same valid age limit has not been added to the Elekta imperial data item which corresponds (eg; weight (lb) ), the calculated result which has no valid age will still be considered valid and could be used in drug calculations. If a stale dose calculation is used in dose calculations the dose may be incorrect. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/12/2015 |