Detail of Recall

Type of Product:Device
Medsafe Reference:19557
Brand Name:Elekta Mosaiq 2.5
Model:Mosaiq
Affected:
Software version:2.5 and higher
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Impac Medical Systems Inc
Issue:Incorrect drug dosage due to "Age Limit" data item issue:
When there is no valid (unexpired) patient weight entered on the same date or prior to a serum creatinine value being populated, a stale Creatinine Clearance (CrCl) value is being used in Area under the Curve (AUC) dose calculations.
Incorrect Drug Dosage due to patient weight data item issue:
If there is a valid age limit added to an Elekta metric data item (eg; weight (kg)) and the same valid age limit has not been added to the Elekta imperial data item which corresponds (eg; weight (lb) ), the calculated result which has no valid age will still be considered valid and could be used in drug calculations.
If a stale dose calculation is used in dose calculations the dose may be incorrect.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/12/2015