Type of Product:Device
Medsafe Reference:19552
Brand Name:Synthes Kirschner Wire
Model:Part No: 02.113.001
Affected:All Lot numbers
Software version:
Recalling Organisation:Johnson & Johnson (NZ) Limited
13a Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Synthes GmbH
Issue:The Kirschner wire is made from Co35Ni35Cr20Mo10 (MP35N) which is cobalt based alloy containing 35% nickel. However, the label and technique guide incorrectly indicate that the devices are made of stainless steel.
The product is shipped with an insert which references the potential risk of allergy/hypersensitivity reactions, but there is no specific information on the label or in the technique guide with regards to the percentage of nickel content in the devices.
Please note: - Product shipped in "kits" in New Zealand are not in packaging and therefore not labelled
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/12/2015