Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19546 |
| Brand Name: | Beckman Coulter Unicel DxH 800 Cellular Analysis System |
| Model: | 629029, B24465, B24802 |
| Affected: | |
| Software version: | 3.0.2.0 |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Brand Name: | Beckman Coulter Unicel DxH SMS Cellular Analysis System |
| Model: | 775222 |
| Affected: | |
| Software version: | 3.0.2.0 |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Brand Name: | Beckman Coulter Unicel DxH 600 Cellular Analysis System |
| Model: | B23858 |
| Affected: | |
| Software version: | 1.1.1.0 |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | During internal testing, Beckman Coulter determined that the software for specific systems allows the creation of multiple orders with the same Specimen Identification (ID) but different Patient Identification when manually editing pending orders at the System manager. The issue does not occur for edits of released results or for test orders requested through host transmission. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/12/2015 |