Detail of Recall

Type of Product:Device
Medsafe Reference:19532
Brand Name:Covidien Endo GIA Ultra Universal Stapler Handles
Model:EGIAUSHORT, EGIAUSTND, EGIAUXL
Affected:Multiple lot numbers
Software version:
Recalling Organisation:Covidien NZ Ltd
Level 3, Building 5, Central Park Corporate Centre
666 Great South Road
Penrose
AUCKLAND 1051
Contact Information:0508 489 264
Manufacturer:Covidien LLC
Issue:A recall is being conducted following customer reports of instruments failing to fire or partially firing and reports of the instrument articulating lever disengaging during use. If the instrument fails to fire or partially fires, or if the articulating lever disengages, the surgeon may be required to replace the device to continue the procedure. No patient injuries related to these issues have been reported. The complaint rate for the instrument failing to fire or partially firing for the affected lots is 0.04% and the complaint rate for the articulating lever disengaging is 0.006%.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/12/2015