Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19509 |
| Brand Name: | BC Access Ostase Calibration Kit and QC Kit |
| Model: | 37305 (Calibrator), 37309 (QC) |
| Affected: | 528936 & 528938 (calibrator); 528937 & 528939 (QC) |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | The Access Ostase Calibrator and Access Ostase QC lots listed above do not meet their labeled expiration dates when stored after initial use (open vial stability). When stored unopened, calibrators and quality controls are stable until the expiration date on the vials. Through internal testing, Beckman Coulter has determined that the above lots are stable at 2 to 10°C for 58 days after initial use. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/11/2015 |