Type of Product:Device
Medsafe Reference:19482
Brand Name:Siemens syngo RT Therapist software
Model:8162815
Affected:
Software version:4.3.138, 4.3.SP1, 4.3.1_AR1, 4.3.1_MR2
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Brand Name:Siemens syngo RT Oncologist software
Model:7356335
Affected:
Software version:4.3.138, 4.3.SP1, 4.3.1_AR1, 4.3.1_MR2
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:The issue occurs if one or more treatment plans with at least 2 isocenters based on one planning CT are used, and when a cone beam CT has been acquired for each isocenter.
The system supports the workflow correctly until the user performs a manual change to another isocenter. At this point the visualization of the overlying images in Adaptive Targeting might be incorrect. If in this case the user performs a manual registration of the overlying images, this may result in an incorrectly calculated offset in Adaptive Targeting.
Applying this offset to the actual patient position can lead to delivering dose to the wrong location and therefore to patient mistreatment.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/11/2015