Detail of Recall

Type of Product:Device
Medsafe Reference:19470
Brand Name:Siemens syngo X-Workplace
Model:105025522 and 10281299
Affected:Software VD10E
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:The issue is caused by a possible incorrect (too small) measurement in the multi-planar reformattings (MPRs) of a volume, which was acquired with a CT scanner with a tilted gantry. In this situation the tabcard "4D" will show incorrect (too small) length measurements. This might potentially result in selecting a device with the wrong size, which then needs to get exchanged.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/11/2015