Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19463
Brand Name:Draeger Babylog VN500 Ventilator with optional PS500
Model:VN500
Affected:With optional PS500 installed
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Brand Name:Draeger Infinity V500 with optional PS500
Model:Infinity V500
Affected:With optional PS500 installed
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Brand Name:Draeger Evita V300 with optional PS500
Model:Evita V300
Affected:With optional PS500 installed
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Issue:Draeger has become aware of cases in which the battery capacity of the optional PS500 of an IACS Workstation Critical Care (Evita Infinity V500, Evita V300 and Babylog VN500 ventilators) was unexpectedly shortened. The batteries were discharged early, even though the device was indicating an adequate battery state of charge. The devices generate a "Battery Low" alarm, followed by "Battery Flat". In exceptional cases, the "Battery Flat" alarm might be displayed too late, or not at all. After complete discharge of the batteries, the device correctly alerts to failure of the power supply by the auxiliary alarm (piezo, power failure alarm) in all cases.
Current analysis of the battery data available indicates that an operating time of at least 20 minutes can normally be maintained after the device is disconnected from the power supply, which sufficiently bridges the gap in the event of a mains power failure until the emergency power supply can be activated.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/11/2015