Detail of Recall

Type of Product:Device
Medsafe Reference:19412
Brand Name:Biomet 3i Provide Protection Caps
Model:PPC484, PPC654, PPC655
Affected:PAK4140, PAK4155, PAK4240, PAK4255, PAK4340, PAK4355, PAK4440, PAK4455, PAK5140, PAK5155, PAK5240, PAK5255, PAK5340, PAK5355, PAK5440, PAK5455, PAK6140, PAK6155, PAK6240, PAK6255, PAK6340, PAK6355, PAK6440, PAK6455
Software version:
Recalling Organisation:ZimVie Australia Pty Ltd (NZ)
16 Huapai Street
Onehunga
Auckland 1061
Contact Information:0508 122 221
Manufacturer:Implant Innovations Inc
Issue:During a review of the Provide® dental restorative product's design history file, inadequate biocompatibility testing was discovered for the Provide Protection Caps (single-use, non-implantable, temporary devices). Subsequent testing was performed and the results indicated a cytotoxic response.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional
Date Commenced:20/11/2015