Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19395
Brand Name:Siemens Artis systems with large display
Model:10094135, 10094137, 10094141, 10094142, 10280959, 10848280, 10848281, 10848282, 10848283, 10848354, 10848355
Affected:Component lots delivered between 18-Dec-2014 and 9-Mar-2015
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue: A loss of video signal could potentially occur in a large display bypass module. This affects displays delivered from 18.12.2014 to 09.03.2015 and is due to improper soldering in limited components of a specific production lot
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/11/2015