Detail of Recall

Type of Product:Device
Medsafe Reference:19362
Brand Name:Molnlycke BARRIER Easywarm and Easywarm+ Blankets
Model:629900 and 629910
Affected:All lots in dating
Software version:
Recalling Organisation:Molnlycke Health Care Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Manukau 2022
Contact Information:0800 005 231
Manufacturer:Molnlycke Health Care AB
Issue:Testing has indicated that Easywarm and Easywarm+ blankets that are more than two years old, do not warm to the specified level of 10 hours. The manufacturer has reduced the shelf life for the blankets from 3 years to 2 years.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/11/2015