Detail of Recall

Type of Product:Device
Medsafe Reference:19357
Brand Name:Philips HeartStart MRx Defibrillator/Monitor
Model:M3535A and M3536A
Affected:
Software version:F.03.06 and T.00.05 and earlier
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:MRx model M3535A with software version F.03.06 and earlier, and model M3536A with version T.00.05 and earlier may stop the automated Ready-For-Use (RFU) test in an abnormal state when the device is turned off. If this occurs, the RFU indicator window displays a pattern that is not documented in the Instructions For Use or Service Manual. If this pattern appears and the device is turned on, it will revert the RFU back to an hourglass, indication that it is fully functional in all monitoring and therapy modes. There is no clinical risk if the device that has displayed this RFU pattern is turned on and put into service.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/11/2015