Detail of Recall

Type of Product:Device
Medsafe Reference:19356
Brand Name:Maquet HCU30 Heater Cooler Unit
Model:Article Number 704631 / 70103.4642
Affected:Selected serial numbers
Software version:
Recalling Organisation:Maquet Australia Pty Ltd (NZ)
12G Andromeda Cres East Tamaki
Auckland 2013
Contact Information:021 874 882
Manufacturer:Maquet Cardio Pulmonary GmbH
Issue:MAQUET Cardiopulmonary has received complaints associated with a failure of the power supply board of the HCU 30. An internal investigation showed that a drop of more than 10% voltage can cause an overheating and subsequent failing of the NTC (Negative Temperature Coefficient) R56 thermistor of the power supply board. This failure typically occurs when the compressor is switched on, i.e. before patient connection. A defective thermistor results in an inoperable compressor or in an inoperable HCU 30. There are no complaints reported associated with a patient injury or death.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/11/2015