Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19345 |
| Brand Name: | Carl Zeiss IOLMaster |
| Model: | 000000-1692-983 |
| Affected: | |
| Software version: | 7.5.2, 7.7.2 and 7.7.3 |
| Recalling Organisation: | Carl Zeiss (NZ) Ltd 22 Fremlin Place Avondale Auckland 1026 |
| Contact Information: | 09 838 5626 |
| Manufacturer: | Carl Zeiss Meditec AG |
| Issue: | Due to a software defect IOL calculation printouts and exported reports can contain wrong IOL power data (mix-up of lens types; calculated IOL power does not fit to stated lens type) if a specific and unusual operator sequence is followed and the healthcare provider does not recognize that the second calculation is identical to the first although a different lens type was selected. This could lead to order and implantation of IOLs with wrong lens power and to (correctable) impairment of patient's visual acuity. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/11/2015 |