Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19334
Brand Name:Hospira Sapphire Infusion Pump
Model:
Affected:Software version 11.00 or lower
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Q Core Medical Ltd
Issue:Under specfic circumstances the Sapphire pump may not detect an occlusion.
The resultant potential delay in therapy may happen if ALL of the following conditions occur:
1. The clamp is left closed at the "start" of the treatment, and
2. The treatment is set at Epidural mode, and
3. The programmed treatment is set to "bolus only" i.e the basal rate =0.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/12/2015