Detail of Recall

Type of Product:Device
Medsafe Reference:19326
Brand Name:Roche Clinical Chemistry tests using NAD(H) or NADP(H) reaction
Model:Products Affected: CKMBL, Creatine Kinase-MB, GLDH3, Gen. 3, NH3 Unisys, Ammonia, NH3L, Ammonia
Affected:All lots
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:Temozolomide may cause interference in assays using NAD(H) or NADP(H). This has been confirmed in assays for GLDH, CKMB and NH3, depending on the analyser used. It has been determined that the interference occurs at either very low analyte, or too high drug concentrations.
These internal investigations were carried out with the therapeutic peak plasma concentration of 0.013 mg/mL Temozolomide.
Higher drug concentrations were not systematically tested, because no literature reference for plasma concentrations > 0.013 mg/mL was found.
The maximum bias for assays which showed interference at 0.013 mg/mL Temozolomide is shown below:
Parameter % bias
GLDH + 62%
CKMB - 31%
NH3 + 36%
This interference may results in falsely high or low results.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/11/2015