Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19312
Brand Name:Siemens N Latex Beta2-Microglobulin
Model:OQWU155
Affected:44862, 44902, 44968, 45192, 45229, 45252, 45291, 45329, 45342, 45366
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Products GmbH
Issue:Siemens Healthcare Diagnostics has received complaints from customers not being able to establish a calibration curve or to validate an established calibration curve with the respective controls when using one of the affected lots of N Latex Beta2-Microglobulin.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/10/2015