Detail of Recall

Type of Product:Device
Medsafe Reference:19310
Brand Name:Roche Benchmark Ultra clinical chemistry analysers
Model:05342716001
Affected:Software Versions VSS 11.9 or higher
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Brand Name:Roche Discovery Ultra clinical chemistry analyser
Model:05987750001
Affected:Software versions VSS 11.9 or higher
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:Internal investigations by Roche Diagnostics have uncovered the potential for incorrect slide staining (IHC/ISH) to occur on the BenchMark ULTRA instrument or DISCOVERY ULTRA instrument running Ventana System Software (VSS) 11.9 or higher, due to the possibility of a duplicate keycode error.
The embedded software for the barcode reader has an identified issue, which on presentation of this error may result in two slides erroneously receiving the same stain protocol regardless of what should have been indicated in the software for that keycode. The barcode reader fails to purge previously read barcode information and applies the same protocol details to a subsequent slide.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/10/2015