Detail of Recall

Type of Product:Device
Medsafe Reference:19303
Brand Name:Siemens Advia Chemistry XPT Systems
Model:SMN 11127538
Affected:Software version 1.0.3
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens has identified a number of issues with the ADVIA Chemistry XPT system with version 1.0.3 that may affect the day to day behaviours and/or workflow of the system:
* Auto start-up failing
* Calibration interval resets when a Reagent Blank (RBL) is run
* Control Definition screen assumes that the range is defined is 2 Std Deviations
* Laboratory Information System (LIS) communciation / Laboratory Automation (LAS) issue
* Printer driver resets
* ISE calibration ranges are too conservative for Urine Sodium (Na)
* Archiving and deletion may fail
* Workstation services may restart
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/10/2015