Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19282
Brand Name:BC Synchron Systems Enzymatic Creatininie CR-E
Model:A60298
Affected:All lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Brand Name:BC Synchron Systems Uric Acid URIC reagent
Model:442785
Affected:All lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:Beckman Coulter has confirmed that N-Acetyl Cysteine (NAC) used in the treatment of acetaminophen overdose may cause assay interference when administered in therapeutic concentrations.
The anti-oxidant properties of NAC can interfere with assays employing the Trinder chemical pathways.
The degree of NAC interference has been demonstrated to vary depending on the analyte. Of the Trinder chemical pathway assays, only Enzymatic Creatinine (CR-E) and Uric Acid (URIC) are affected.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/10/2015