Detail of Recall

Type of Product:Device
Medsafe Reference:19259
Brand Name:Roche Virtuoso software
Model:07299303001 (V5.4), 07072984001 (V5.5)
Affected:
Software version:5.4 and 5.5
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:The iScan Coreo or VENTANA iScan HT scans slides at 4x, 10x, 20x, or 40x magnifications. For images scanned at 4x, 10x, and 40x magnification in v5.4 and v5.5, the linear and area measurement tools display a value that is 1/5, 1/2 and 2x the actual measurement, respectively. The measurement will be displayed incorrectly regardless of what magnification the image is viewed at.
Customers who have used the measurement tool for slides that were scanned at 4x, 10x, or 40x to assist with a diagnosis, need to be aware that an incorrect estimation of clinical stage could lead to improper treatment.
For images that are scanned at 20x magnification, the measurement tools work properly in all versions of Virtuoso including versions 5.4 and 5.5.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/10/2015