Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19252 |
| Brand Name: | Siemens Enzygnost Anti-VZV/IgG |
| Model: | OWLT15 |
| Affected: | 44218, 44309, 44565, 45077 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Products GmbH |
| Issue: | Siemens Healthcare Diagnostics Products has confirmed that selected Enzygnost Anti-VZV/IgG lots can show a higher variation in homogeneity. This can lead to invalid runs by not meeting the run validation criterion for the Reference P/N at the start and end of a series. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/10/2015 |