Detail of Recall

Type of Product:Device
Medsafe Reference:19249
Brand Name:GE Giraffe Shuttle Uninterruptible Power Supply
Model:Shuttle
Affected:Serial numbers MAAU61000 or lower
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Ohmeda Medical Inc
Issue:GE Healthcare has recently become aware of a potential issue with the shortened battery life. Shortened battery life can lead to a loss of power to devices plugged into the Shuttle. Power loss during transport due to this battery issue may lead to a minor delay in the transport time.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/10/2015