Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19243
Brand Name:Ethicon Harmonic Ace+ 7 Laparoscopic Shears
Model:HARH23, HARH231, HARH36, HARH361, HARH45 and HARH451
Affected:Expiration dates prior to 2020-08
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Ethicon Endo-Surgery LLC
Issue:Ethicon, Inc. have determined that under extreme handling circumstances, damage, such as a hole, can occur in the Tyvek® lid due to an interaction between the Advanced Haemostasis (green) button and the Tyvek® lid which could compromise device sterility. Ethicon has received no reports of Adverse Events for this issue.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/10/2015
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /