Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19230 |
| Brand Name: | Hospira Plum 360 Infusion pump |
| Model: | 30010-27 |
| Affected: | |
| Software version: | 15.01.XX.009 |
| Recalling Organisation: | Hospira NZ Ltd 58 Richard Pearse Drive Airport Oaks Mangere Auckland 2022 |
| Contact Information: | 04 384 7463 |
| Manufacturer: | Hospira Inc |
| Issue: | Hospira has determined if "Yes" is selected at the "New Patient" screen, and the program was not cleared prior to powering off the infuser, the Distal Occlusion Pressure setting is assigned a value of 15 PSI (775 mmHg) instead of the default value. In addition, the User Interface will incorrectly display the default Distal Occlusion Pressure value, which is set at 6 PSI (310 mmHg) from the manufacturer. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/10/2015 |