Detail of Recall

Type of Product:Device
Medsafe Reference:19230
Brand Name:Hospira Plum 360 Infusion pump
Model:30010-27
Affected:
Software version:15.01.XX.009
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Hospira Inc
Issue:Hospira has determined if "Yes" is selected at the "New Patient" screen, and the program was not cleared prior to powering off the infuser, the Distal Occlusion Pressure setting is assigned a value of 15 PSI (775 mmHg) instead of the default value. In addition, the User Interface will incorrectly display the default Distal Occlusion Pressure value, which is set at 6 PSI (310 mmHg) from the manufacturer.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/10/2015