Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19220
Brand Name:Stryker Vitallium Femoral Heads with PCA taper, LFIT V40 taper or V40 taper
Model:6260-5-028, 6260-5-032, 6260-5-132, 6260-5-232, 6260-5-328, 6260-5-332, 6260-5-428, 6260-5-432, 6260-9-028, 6260-9-032, 6260-9-132, 6260-9-232, 6260-9-328, 6260-9-332, 6260-9-428, 6280-0-128, 6280-0-132 and 6280-0-228
Affected:Manufactured between 7-Jul-2014 and 15-Aug-2014
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Howmedica Osteonics
Issue:Stryker has received four customer complaints for V40 LFIT Vitallium femoral heads (manufactured 7-Jul-2014 and 15-Aug-2014) reporting, at the time of surgery the femoral head could not be assembled with its corresponding V40 stem trunnion. In each case a new V40 LFIT Vitallium femoral head was opened and used. No other adverse consequences or delays to surgery were reported for any of these complaints.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/10/2015
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /