Detail of Recall

Type of Product:Device
Medsafe Reference:19212
Brand Name:Roche PreciControl ClinChem Multi 1
Model:PCCC1
Affected:174787 & 174794
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Brand Name:Roche PreciControl ClinChem Multi 2
Model:PCCC2
Affected:174801
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:Roche has received a number of complaints that the recovery of C-Reactive Protein Gen. 3 (CRPL3) was too high in PreciControl ClinChem Multi 1 (PCCC1), and especially in PreciControl ClinChem Multi 2 (PCCC2), seen mainly on the cobas c 501 module.
Roche Diagnostics has investigated and determined that the last three reagent lots showed higher recovery for all control materials. To resolve this issue a revised target value assignment procedure was necessary.
Target values for currently available PCCC1 and PCCC2 control lots have been adjusted as per the tables given in the Customer Letter.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/10/2015