Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19204 |
| Brand Name: | Neoss Guide - Drill hubs I to V included in Neoss Guide Kit |
| Model: | 51153 |
| Affected: | 15182, 15586, 16030 |
| Software version: | |
| Recalling Organisation: | Neoss Australia Pty Ltd (NZ) 33 Kowhai Street Hamilton Lake Hamilton 3204 |
| Contact Information: | 021 464 344 |
| Manufacturer: | Neoss Ltd |
| Issue: | Specific batches of Drill-Hub I may not be functioning as intended (drill does not latch completely in the Drill-Hub I). The upper part of the bore in the Drill-Hub I is slightly smaller which results in the drill not being seated correctly which means that this Drill-Hub will drill deeper than intended. This could lead to permanent damage of the alveolar nerve and cause permanent loss of sense in parts of the oral cavity or chin or even partial loss of speech. There is also a risk of drill loosening or not rotating as expected. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/9/2015 |