Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19187
Brand Name:Medifab Shuttle Discovery System
Model:2006-0010-000
Affected:JS-06-14-DS-**, JS-15-14-DS-**, JS-18-14-DS-**, JS-23-14-DS-**, JS-45-14-DS-**, JS-46-14-DS-**, ML-8-15-DS-**, ML-13-15-DS-**, ML-16-15-DS-**, ML-17-15-DS-**, ML-18-15-DS-**, 61427, 61738
Software version:
Recalling Organisation:Medifab Limited
1 Pereita Drive
Rolleston 7675
Contact Information:0800 543 343
Manufacturer:Medifab Limited
Issue:The manufacturer has identified two potential issues relating to the backrest and the tilt in space locking plunger should the vehicle used to transport the end user in the Discovery Buggy be involved in an accident.
1/ There is potential for the S Tubes to come out of the mounting block
2/ There is potential that the Discovery seat may tilt back due to the tilt in space plungers not being engaged in the plastic slides
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Public Recall
Date Commenced:6/10/2015