Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19187 |
| Brand Name: | Medifab Shuttle Discovery System |
| Model: | 2006-0010-000 |
| Affected: | JS-06-14-DS-**, JS-15-14-DS-**, JS-18-14-DS-**, JS-23-14-DS-**, JS-45-14-DS-**, JS-46-14-DS-**, ML-8-15-DS-**, ML-13-15-DS-**, ML-16-15-DS-**, ML-17-15-DS-**, ML-18-15-DS-**, 61427, 61738 |
| Software version: | |
| Recalling Organisation: | Medifab Limited 1 Pereita Drive Rolleston 7675 |
| Contact Information: | 0800 543 343 |
| Manufacturer: | Medifab Limited |
| Issue: | The manufacturer has identified two potential issues relating to the backrest and the tilt in space locking plunger should the vehicle used to transport the end user in the Discovery Buggy be involved in an accident. 1/ There is potential for the S Tubes to come out of the mounting block 2/ There is potential that the Discovery seat may tilt back due to the tilt in space plungers not being engaged in the plastic slides |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Public Recall |
| Date Commenced: | 6/10/2015 |