Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19175
Brand Name:Terumo Advanced Perfusion System 1
Model:801764
Affected:All units
Software version:
Recalling Organisation:Terumo Australia Pty Ltd (NZ)
c/- Healthcare Logistics
56 Carrington Road
Mount Albert
Auckland 1025
Contact Information:0800 66 77 57
Manufacturer:Terumo Cardiovascular Systems Corp
Issue:Terumo Cardiovascular Systems (Terumo CVS) has become aware that under certain circumstances, a Terumo® Advanced Perfusion System 1 centrifugal pump "Service Pump" message may display due to user induced motor movement.
In this instance, the Terumo System 1 will recognize the movement (as it is designed to do) and display a "Service Pump" message. The user might interpret this message as a pump that requires service rather than a message displayed due to pump movement while not activated.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/9/2015