Detail of Recall

Type of Product:Device
Medsafe Reference:19174
Brand Name:Dako EnVision FLEX/HPR reagent
Model:K8000 & K8012
Affected:20019097 & 20020768
Software version:
Recalling Organisation:Global Science a Bio-Strategy Company (Bio-Strategy Limited)
241 Bush Road
Albany
Auckland 0632
Contact Information:0800 734 100
Manufacturer:DakoCytomation
Issue:Dako has determined that there is a defect in one lot of a buffer used in the manufacturing of some corresponding lots of the EnVision™ FLEX/HRP visualization reagent. Specifically, some instances of moderate-to-strong non-specific staining and overly-strong specific staining have been observed when this visualization reagent was used with four antibodies against the biomarkers Epstein-Barr Virus (EBV), MUM-1, Wilms' Tumor1(WT1) Protein and Smooth Muscle Actin (SMA). Negative control antibodies in the same run were negative, and did not show the non-specific staining.

15/01/2016: Subsequent to the earlier letter the remaining Dako FLEX Ready-to use antibodies were tested using the affected EnVision FLEX/HRP visualization reagent . Two additional antibodies wereidentified as having a remote potential to affect results. Update overview letter sent.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/9/2015