Detail of Recall

Type of Product:Device
Medsafe Reference:19165
Brand Name:Olympus GF-UCT180 Ultrasonic Gastrovideoscope
Model:GF-UCT180
Affected:All serial numbers
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:There is a risk of patient-to-patient infection if the device is not cleaned appropriately.
Key changes in the new Instruction for Use are highlighted below:
* Inclusion of instruction on how to attach the cleaning adapter (MAJ-350) to the instrument channel port of the GF-UCT180
* The balloon applicator (MAJ-675) and the cleaning brush (MAJ-1534) are to be reprocessed for reuse. The MAJ-675 and MAJ-1534 are to be reprocessed in accordance with the cleaning, disinfection, and sterilisation procedures for reusable parts and reprocessing equipment.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/9/2015