Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19158 |
| Brand Name: | Liko Viking Mobile Lifts |
| Model: | Viking M (2040035, 2040015, 2040005), Viking L (2040004), Viking XL or 300 (2040003) |
| Affected: | Multiple serial number ranges |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | Liko AB |
| Issue: | The manufacturer has received 9 complaints where the lift arm drifted down suddenly on certain Viking mobile lifts. These reports have been connected to users who raise the lift arm assembly manually. The lift arm assembly is intended to only be raised by the actuator (lift motor). If the lift arm assembly is manually raised, the actuator can become damaged and get stuck in the highest position. When the actuator is stuck, the lift cannot be raised or lowered by the lift motor. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/10/2015 |