Detail of Recall

Type of Product:Device
Medsafe Reference:19158
Brand Name:Liko Viking Mobile Lifts
Model:Viking M (2040035, 2040015, 2040005), Viking L (2040004), Viking XL or 300 (2040003)
Affected:Multiple serial number ranges
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:Liko AB
Issue:The manufacturer has received 9 complaints where the lift arm drifted down suddenly on certain Viking mobile lifts. These reports have been connected to users who raise the lift arm assembly manually.
The lift arm assembly is intended to only be raised by the actuator (lift motor). If the lift arm assembly is manually raised, the actuator can become damaged and get stuck in the highest position. When the actuator is stuck, the lift cannot be raised or lowered by the lift motor.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/10/2015