Detail of Recall

Type of Product:Device
Medsafe Reference:19151
Brand Name:GE 3.0T MRI Scanners
Model:Discovery MR750w 3.0T, Discovery MR750 3.0T, SIGNA PET/MR 3.0T, SIGNA HDxt 3.0T, SIGNA HDx 3.0T, SIGNA HD 3.0T, SIGNA Excite 3.0T, SIGNA 3.0T
Affected:All units
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Issue:GE Healthcare has recently become aware of a potential safety issue when performing head or neck scans on 3.0T MR scanners. The currently displayed Head Specific Absorption Rate (SAR) values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/9/2015