Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19096 |
| Brand Name: | Terumo Spectra Optia Apheresis System |
| Model: | Optia |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | The Advisory Group Ltd Level 9 55 Shortland St Auckland 1010 |
| Contact Information: | |
| Manufacturer: | Terumo BCT Inc |
| Issue: | The manufacturer has received reports of data entry errors that cause the system to calculate a patient total blood volume (TBV) that is abnormal for that specific patient. If the operator does not enter correct data the patient may receive excess anticoagulant (AC) or the outcome of the procedure may be affected. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/9/2015 |