Detail of Recall

Type of Product:Device
Medsafe Reference:19089
Brand Name:Siemens RapidPoint 500 System, Lactate cartridge
Model:10491449. 10491448, 10491447 and 10844813
Affected:Cartridge serial numbers 2517102517 through 2519000012
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:During the manufacturing process the Sodium high calibrator value was incorrectly assigned for the RAPIDPoint 500 measurement cartridge serial numbers indicated above causing a negative bias on both QC and patient samplers.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/9/2015