Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19083
Brand Name:Boston Scientific Watchman Left Atrial Appendage Closure Device with Delivery System
Model:All models
Affected:Selected lots
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Boston Scientific Corporation
Brand Name:Boston Scientific Watchman Access System Access Sheath with Dilator
Model:All models
Affected:Selected lots
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Boston Scientific Corporation
Issue:The manufacturer has determined cross-threading of the hemostasis valve may occur if the valve is tightened with the dilator in place, potentially preventing subsequent sealing of the valve when desired.
The manufacturer has issued a recall for product correction to reinforce existing Directions for Use (DFU) and provide further guidance regarding the correct use of the hemostasis valve in order to avoid cross-threading and to securely seal the valve, minimizing the potential for undesirable blood leakage.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/9/2015