Detail of Recall

Type of Product:Device
Medsafe Reference:19061
Brand Name:Iris iChem Velocity Chemistry Strips
Model:800-7204
Affected:Lot: 7204081A
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Iris International
Issue:The manufacturer has become aware of a manufacturing defect where a glucose pad was placed in the location intended for the protein pad on a subset of vials numbered 14996 through 15859 (with the exception of vials 15315 and 15316 within that series) within lot number 7204081A. The hazarous situation is false negative or false positive urine protein for patient samples released to the physician.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/10/2015