Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19058 |
| Brand Name: | Radiometer ABL90Flex analyser, battery module |
| Model: | 910-311 |
| Affected: | Battery module serial number below 910-311P0009Nxxx |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | There is potential for battery modules to overheat caused by continued charging of a collapsed battery cell. This can occur in batteries from production runs below 910-311P0009Nxxx. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/9/2015 |