Detail of Recall

Type of Product:Device
Medsafe Reference:19058
Brand Name:Radiometer ABL90Flex analyser, battery module
Model:910-311
Affected:Battery module serial number below 910-311P0009Nxxx
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:There is potential for battery modules to overheat caused by continued charging of a collapsed battery cell. This can occur in batteries from production runs below 910-311P0009Nxxx.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/9/2015