Detail of Recall

Type of Product:Device
Medsafe Reference:19032
Brand Name:Siemens Chemistry XPT ISE
Model:10723034
Affected:All units
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens is reminding users that the Ion Selective Electrodes (ISE) must be calibrated after replacement and controls run. If some electrodes are not plugged in, it is possible for the system to produce a discrepant, unflagged, result for sodium, potassium or chloride.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/9/2015