Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19032 |
| Brand Name: | Siemens Chemistry XPT ISE |
| Model: | 10723034 |
| Affected: | All units |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | Siemens is reminding users that the Ion Selective Electrodes (ISE) must be calibrated after replacement and controls run. If some electrodes are not plugged in, it is possible for the system to produce a discrepant, unflagged, result for sodium, potassium or chloride. |
| Recall action type: | Alert |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/9/2015 |