Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19010
Brand Name:CareFusion Alaris Syringe Infusion pump
Model:8110
Affected:Manufactured between Mar-2014 and Sep-2014
Software version:
Recalling Organisation:CareFusion New Zealand 313 Limited
Unit 4 / 110 Mays Road
Onehunga
AUCKLAND 1061
Contact Information:0508 422 734
Manufacturer:CareFusion 303 Inc
Issue:CareFusion has identified an issue with the Alaris Syringe Pump module 8110 because the channel error will cause an audible and visual alarm on the Alaris PC unit and a channel error on the Alaris Syringe. Once cleared on the PCU the Syringe module remains unresponsive to key presses. This error could occur during an infusion causing an interruption of infusion.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/11/2015