Detail of Recall

Type of Product:Device
Medsafe Reference:19005
Brand Name:Volcano s5/s5i/CORE/CORE Mobile Systems with FFR option
Model:Volcano s5 and CORE series
Affected:Software versions 3.3, 3.4 (with and without iFR Scout technology) and v3.4 software kits
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Volcano Corporation
Issue:Volcano Corp. has discovered a very specific workflow scenario where the Aortic pressure (Pa) offset used in the FFR/iFR modalities will be lost or reset to 0 mmHg due to a software error. This issue can lead to an inaccurately calculated FFR or iFR value. Should this occur, a clinician may rely on an inaccurate value in diagnosing a patient's condition.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/9/2015